Normalize; Store Reference Trace - Agilent Technologies PSA Series User Manual

Option 215 for the psa series of spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Source Key and Programming Commands

Normalize

4.5 Normalize

4.5.1 Store Reference Trace

Copies trace 1 into trace 3.
Mode:
Key Path:
Remote Command:
Dependencies and
Couplings:
Example:
58
Spectrum Analysis
Source, Normalize
:TRACe:COPY TRACE1, TRACE3
Parameter is disabled if active mode is not Spectrum Analyzer
or a connection has not been established at the specified IP
address.
A reference trace should be stored prior to turning on trace
normalization either through pressing the Store Ref Trace
softkey or completing an Open/Short calibration.
The stored reference trace (trace 3) is cleared when key is
pressed. If you wish to re-use this calibration data you should
save the trace (File > Save > Type (Trace) > Save Now) and then
recall it following completion of the preset.
TRAC:COPY TRACE1, TRACE3