Show Setup; Model Number - Agilent Technologies PSA Series User Manual

Option 215 for the psa series of spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Source Key and Programming Commands
External Source Configuration

4.1.2 Show Setup

Pressing this immediate action key displays a form showing you the current configuration of the external
source at the specified IP address.
Mode:
Key Path:
Dependencies and
Couplings:
Remote Command
Notes:

4.1.3 Model Number

Information parameter which displays the model number of the signal source connected at the specified IP
Address. The model number displayed will be determined by sending a "*IDN?" to the entered IP address.
The model number is the second field of the "*IDN?" returned value (Format of the *IDN? string is
'<Company Name>, <Model Number>, <Serial Number>, <Firmware Revision>').
Mode:
Key Path:
Remote Command:
Factory Preset:
State Saved:
Range:
Remote Command
Notes:
Example:
48
Spectrum Analysis
Source, Ext Src Config
Parameter is disabled if active mode is not Spectrum Analyzer
or a connection has not been established at the specified IP
address.
No SCPI. Front Panel only.
Spectrum Analysis
Source, Ext Src Config, Show Setup
:SOURce:EXTernal:MODel?
------
No
Uppercase, Numeric
Query Only
SOUR:EXT:MOD?
Chapter 4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Psa e4440aPsa e4443aPsa e4445aPsa e4446aPsa e4448a