6.2. Exchanging Sensors

It is recommended to keep the dew-point sensor with the control unit that it was originally
ordered with. However, if it is necessary to replace the sensor or exchange it for a spare,
there are two steps which need to be taken.
1. Connect to the control unit via the application software (as detailed in the Application
Software section at the end of the manual) and click the 'Enter Sensor Configuration' button
at the Main Options window and enter the password 7316Sens.
2. Enter the 12 character configuration code found on the calibration certificate for the sensor
you are connecting. Connect the new sensor, then navigate to the 'Alarms' screen, and press
the 'Calibrate Optics' button. A DCC will follow, which cannot be cancelled. Do not disconnect
the sensor during this time.