• Opening or altering the device invalidates the guarantee.
• The guarantee does not cover damage caused by improper
handling, a discharged battery, accidents or non-compliance
with the operating instructions.
Please contact Microlife-service.
14. Technical Specifications
Type:
Ear Thermometer IR 200
Measurement
range:
0-100.0 °C / 32-212.0 °F
Resolution:
0.1 °C / °F
Measurement
Body mode:
±0.2 °C, 35.0 ∼ 42.0 °C / ±0.4 °F, 95.0 ∼ 107.6 °F
accuracy
(Laboratory):
±0.3 °C, 34.0 ∼ 34.9 °C and 42.1 °C ∼ 43.0 °C /
±0.5 °F, 93.2 ∼ 94.8 °F and 107.8 ∼ 109.4 °F
Object mode:
±1.0 °C, 0 ∼ 100.0 °C / ±2 °F, 32.0 ∼ 212 °F
Display:
Liquid Crystal Display, 4 digits plus special icons
Acoustic:
The unit is turned ON and ready for the measure-
ment: 1 short beep.
Complete the measurement: 1 long beep.
System error or malfunction: 3 short beeps.
Fever alarm: 10 short beeps.
Memory:
30 readings recall in the memory mode with a
record of both time and date.
Backlight:
The display light will be GREEN for 1 second,
when the unit is turned ON.
The display light will be GREEN for 5 seconds,
when a measurement is completed with a reading
less than 37.5 °C / 99.5 °F.
The display light will be RED for 5 seconds, when
a measurement is completed with a reading equal
to or higher than 37.5 °C / 99.5 °F.
Operating
10 - 40 °C / 50.0 - 104 °F
conditions:
15-95 % relative maximum humidity
Storage
-25 - +55 °C / -13 - +131 °F
conditions:
15-95 % relative maximum humidity
Automatic
Approx. 1 minute after last measurement has been
Switch-off:
taken.
Battery:
1 x CR2032 battery 3V
IR 200
Battery
lifetime:
approx. 1000 measurements (using a new battery)
Dimensions: 150 x 43 x 60 mm
Weight:
60 g (with battery), 58 g (w/o battery)
IP Class:
IP22
Reference to
EN 12470-5; ASTM E1965;
standards:
IEC 60601-1; IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected
service life:
5 years or 12000 measurements
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial technical
inspection is recommended for professional users. Please
observe the applicable disposal regulations.
15. www.microlife.com
Detailed user information about our thermometers and blood pressure
monitors as well as services can be found at www.microlife.com.
EN
7