Before switching on, check whether the centrifuge is connected to power supply
correctly.
Before using check whether the device is correctly intalled.

Current protection

The centrifuge is equipped with current protection (safety fuse). Fuse is situated in
the plug-in socket unit at back wall of the centrifuge.
Fig.1 Plug-in socket unit
Do not position the centrifuge so that it is difficult to operate the power switch
Supply voltage given on the rating plate has to be consistent with local supply
voltage. LABNET INTERNATIONAL laboratory centrifuges are 1st safety class
devices and they are provided with the three-core cable with the plug resistant
to dynamic loadings. Mains socket shall be provided with the safety pin
- protective earth (PE).
It is recommended to install emergency cut-out that shall be located far from
the centrifuge, near the exit or beyond the room.
Safety fuse
© 2018 Corning Incorporated
9