Emergency lid release
In case of e.g. mains failure it is possible to open cover manually. On the right side is
plug, which should be unscrew (via key for emergency lid release 18640 basic
accessories). Then, one should pull the plug.
point of emergency
lid releasing
It is not allowed to emergency lid releasing when rotor is running!
One must be sure that rotor is not in the motion (use inspection glass).
Screen failure
If the information displayed on the screen disappears and there is no display backlight,
press the STOP button twice to stop the centrifuging. Then, make sure to look through
the viewfinder into the centrifuge chamber that the rotor has stopped rotating and
turn off the power supply with the power switch.
For reasons of safety, do not use the lid emergency opening. 5 minutes after stopping
the rotor turn on the power supply with the power switch.