Maintenance
7. Reassemble pipette top.
Attach pipette top to the
main housing. Ensure flex
cable is behind the top of
the pipette trigger and the
trigger contacts the limit
switch properly (see Figure
F.3.1Ae). Fit the edges of the
pipette top snugly within the
edges along the top of the
pipette housing. Ensure that
the cables to the power jack
are inside the main housing
well below the edges to avoid
pinching the cables. Hold
assembly together firmly
while resecuring the two
screws.
8. Charge battery. Charge the
new battery before using the pipette
(see Section F.4, Recharging Batteries).
F.3.1B Installing/Replacing Multichannel Electronic Pipette Batteries
The Matrix Electronic pipette is shipped with Nickel Metal Hydride batteries
packaged separately. Install the batteries and recharge them for 14-16 hours.
Use the following procedure to install the batteries, then continue to Section
F.4, Recharging Batteries.
46
Figure F.3.1Ba: Removing Battery Case
Figure F.3.1Ae:
Assembling Pipette Top