Programming and Operation
4. Dispense liquid. Position the tips over the first column of wells in the
microtiter plate, and press the trigger to dispense 30 µl. At the completion
of the dispense step, the pipette emits a single beep tone (if the beep tone
is set to ON1).
5. Repeat dispense. Repeat the previous step for the rest of the columns in
the microtiter plate. At the completion of the eighth dispense, the pipette
emits a double beep tone (if the beep tone is on, otherwise it emits a single
beep tone) and displays the "Purge" prompt. There is insufficient volume
for another 30 µl dispense, and the pipette signals for a purge step.
-or-
Paced dispense. Steps 4 and 5 can also be performed as a paced
dispense. Instead of pressing the trigger for each dispense step, press
and hold the trigger to automatically dispense each volume. Immediately
after each dispense step, the pipette emits a single beep tone (if the
beep tone is on), prompting you to prepare for the next dispense step.
To stop a paced dispense, release the trigger before the end of the last
dispense step.
Note: For paced dispensing of small volumes, set the pace to SLOW. Ensure
that you have fully released the trigger before the end of the last dispense step
to prevent an additional, unwanted dispense step.
6. Discard remaining liquid. Position the pipette tips over a waste container
and press the trigger to purge the remaining liquid. The pipette emits a
triple beep tone to signal the end of the program, then returns to the first
step of the program.
7. Eject tips. Press the ejector trigger to eject the used tips.
22