Note: If the Cross Tube orientation between the cylinder and Tip Fittings is
incorrect, performance of the pipette will be impaired.
3. Carefully insert the flat end of the motor screw into the hole of the motor
coupling. This can be accomplished by holding the piston/cylinder
assembly at a slight angle while inserting the motor screw into the hole
of the motor coupling. Slowly spin the motor screw until the flat end of
the motor screw is fully engaged into the motor coupling hole (see Figure
F.2.3e). The cylinder can now be seated between the plastic ribs in the
main housing.
NOTE: The lot number on the face of the cylinder should be facing toward
you and the painted section of the piston backplate should be facing the
photosensor. The painted section is used by the photosensor to track the
location of the pistons
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
3
4
6
7
1
2
3
10
11
1
2
3
10
11
1
2
3
4
5
1
3
4
6
7
1
3
4
5
7
Table F.2.3b: Cylinder/Cross Tube Positions
.
Tip Fitting Positions
6
7
8
9
10
Cylinder Positions
6
7
8
9
10
6
7
8
9
10
9
10
12
12
6
7
8
8
Maintenance
11
12
13
14
15
11
12
13
14
15
11
12
16
16
41