Maintenance
Figure F.2.1b: Removing Ejector
Trigger and Ejector Comb
36
Figure F.2.1c: Removing Piston/Cylinder
and Manifold Assemblies
4. Remove plastic ejector
trigger with spring and
slide the aluminum ejector
comb off of the tip fittings
(see Figure F.2.1b). On 16
channel models, reverse
this process and remove
comb first, then trigger.
5. Remove the piston/
cylinder assembly by
pulling the motor screw
down from the motor shaft
coupling and out of the
main housing. Remove
the manifold assembly by
firmly lifting it out of the
main housing support ribs
(see Figure F.2.1c).