Thermo Scientific Matrix 1029 User Manual page 48

Electronic pipette
Hide thumbs Also See for Matrix 1029:
Table of Contents

Advertisement

6. Hold the ejector comb in place while inserting the ejector trigger (see
Figure F.2.3h). Place the end of the ejector spring against the plastic stop
and compress the spring while positioning the ejector trigger over the
aluminum ejector comb (the ribs on the ejector trigger should mate with the
slots on the aluminum ejector comb). For 16-channel models, install the
trigger prior to connecting the ejector comb.
NOTE: Be careful not to pinch the cross tubes when inserting the ejector
trigger.
Test the ejector trigger to ensure that the ejector comb moves down when
the trigger is pressed (hold the ejector comb firmly in place while testing
the trigger).
7. Place the cover housing onto the main housing and fasten the
three screws.
8. Re-attach the color ring.
9. Test pipette for smooth operation. (Reassemble and repeat steps
3-7 if neccessary.)
The pipette is now ready for use.
Figure F.2.3h: Inserting Ejector Trigger
Maintenance
43