Prepare The Logic Analyzer For Testing - Agilent Technologies 16800 Series Service Manual

Portable logic analyzers
Hide thumbs Also See for 16800 Series:
Table of Contents

Advertisement

Prepare the Logic Analyzer for Testing

16800 Series Portable Logic Analyzers Service Guide
1
Record the logic analyzer's model and serial number in
the Performance Test Record (see
work order number (if applicable) and today's date.
Record the test equipment information in the "Test
2
Equipment Used" section of the Performance Test Record.
Turn on the logic analyzer.
3
Connect the keyboard to the rear panel of the logic
a
analysis mainframe.
b
Connect the mouse to the rear panel of the mainframe.
c
Plug in the power cord to the power connector on the
rear panel of the mainframe.
Turn on the main power switch on the mainframe
d
front panel.
While the logic analyzer is booting, observe the boot
dialog for the following:
Ensure all of the installed memory is recognized.
Any error messages.
Interrupt of the boot process with or without error
message.
4
During initialization, check for any failures.
If an error or an interrupt occurs, refer to
"Troubleshooting," starting on page 71.