10. DATA INTERPRETATION
12. Touch <More Options> to advance to the Fall Risk Test Options screen if desired. Here you
can set the Test Trial Time, enter the number of trials and enter the Rest Countdown. You
can also toggle the cursor ON/OFF. Touch <OK> to confirm your selections and return to
the Fall Risk Testing screen.
Use the < > or <M> keys to set the Test Trial time in five-second increments
(during the routine the system will count down from the time setting selected).
To turn the cursor ON/OFF, touch <Cursor> to toggle between choices.
To set the number of trials or rest countdown, touch the appropriate key and then enter
the setting from the keypad displayed.
13. Press <Start> to release the platform (if not static) and activate the Fall Risk Test screen.
14. With the patient ready to begin the test, touch <Collect Data>. The screen will provide a
three-second countdown before beginning the first test trial. The display screen will show
Test Trial Time and Platform Setting to the left of the grid. Trial Number and score are dis-
played to the right of the grid.
NOte: To stop a test in progress at any time and return to the Fall Risk Testing screen with the plat-
form locked, touch <Cancel> or <Stop>.
15. When the first trial is finished, the screen will display "Trial 1 Complete," the platform will
return to the locked position, and the rest countdown will begin for the second trial. Touch
<Collect Data> to begin the second test trial and continue in the same manner to complete
trials two and three.
16. After completing the test, a "Test Complete" message is displayed. Touch <Results> to
advance to the Fall Risk Test Results screen.
17. At the Fall Risk Test Results screen, touch <Print> to automatically generate a printed report
if desired.
18. To save the test data, touch <Save Results> and then touch <OK> in response to the "Save
Results for later reporting or export?" prompt. The system will display "Save Results
Completed" after the results are saved.
19. To return to the Opening Menu from the Fall Risk Test Results screen touch <Home>.
TESTING