Download Print this page

Otto Bock 8E70 Instructions For Use Manual page 42

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 32
If the interference cannot be eliminated or if you do not achieve the expected results by
adjustment or selection of the appropriate control programme, please contact the Ottobock
branch responsible for your country.
CAUTION
Prosthesis use when battery charge level is too low
Injury due to unexpected prosthesis behaviour
Check the current charge level before use and charge the prosthesis if required.
Note that the operating time of the prosthesis may be reduced at low ambient temperatures
or due to ageing of the battery.
Note that the actions/reactions of the terminal device become slower when the battery
voltage gets very low.
Note that only a few grips or actions will still be possible with the terminal device when the
battery voltage gets very low.
A small opening width may indicate a low battery voltage.
4.4 Information on Alignment/Adjustment
CAUTION
Operator errors during the adjustment process with the adjustment software
Injury due to unexpected product behaviour.
Participation in an Ottobock product training course is required prior to initial use. Addition­
al product training courses may become necessary to qualify for software updates.
Transfer the setting changes to the terminal device before you check the settings on the
patient.
Use the online help which is integrated into the software.
CAUTION
Use of unapproved accessories
>
Injury due to product malfunction as a result of reduced resistance to interference.
>
Interference of other electronic devices due to increased emissions.
Only use the product in combination with the accessories, signal converters and cables lis­
ted in the sections "Combination possibilities" (see Page 39), "Scope of delivery" (see
Page 45) and "Accessories" (see Page 45).
CAUTION
Use of unsuitable prosthetic components
Injury due to unexpected product behaviour.
Use the product only in combination with components listed in the section "Combination
possibilities" (see Page 39).
CAUTION
Failure to observe the instructions for use of all prosthesis components used
Injury due to unexpected product behaviour.
Observe the instructions for use of the prosthesis components used.
42

Advertisement

loading

This manual is also suitable for:

8e718e72