6
Repair
6.1
Safety-related tests
A protective conductor test and a leakage current test must be performed
prior to the installation or the hand-over of the unit as well as after repair
work.
See Sections "Protective conductor test [ → 63]" and "Leakage current
test."
6.2
Replacing X-ray tube assembly H1
Required tools
Torx screwdriver (size 10, 15 and 20)
Removal of the X-ray tube assembly
1. Switch the X-ray unit off.
2. Unscrew and remove the old cover (A).
3. Remove the old arm cover (L).
L
6 Repair
6.1 Safety-related tests
81