Dentsply Sirona
Service Manual Heliodent Plus
62 15 102 D3507
D3507.076.01.15.02 03.2018
4.2
Checking the protective ground connections
4.2.1
Test
➢ Check whether all protective ground connections are firmly in place.
4.2.2
Positions
DX1 and installation panel (wall module)
● X200 / Power input clamp
● X500* / Output stage and installation panel
– (A): Screw
* The protective ground wire is only connected to X500 / PE up to
serial No. 26 999, from serial No. 27 000 the protective ground
wire is only screwed onto the screw (A).
4.2 Checking the protective ground connections
4 Maintenance
41