Calibration Test Report
The calibration test mode is used to compare a calibration weight with the calibration test data weighed by
the scale.
This test does not perform calibration.
Step 1
In the weighing mode, press and hold the CAL switch until
1CC
appears, and then release the switch, or, press and
hold the
C
and release the key.
The calibration test mode is not available when the function
setting inf1 0 is selected.
Step 2
Press the ENTER key to display
Step 3
Change the calibration weight value if necessary.
To change the calibration weight value, press the MODE key.
Change the value using the following keys.
Numerical keypad Enter the value.
ENTER key
ZERO key
Step 4
With nothing on the weighing pan, press the ENTER key.
The zero point is measured and the weighing value with the
unit "kg" is displayed for a few seconds. Then, the value of
the calibration weight is displayed.
Step 5
Place a calibration weight of the same value as displayed on
the weighing pan and press the ENTER key to measure it.
The measured value with the unit "kg" is displayed for a few
seconds.
Step 6
end appears.
gmp appears and the calibration test report is output.
Step 7
CC
Step 8
appears again. Remove the weight.
Press the ON/OFF key to turn the display OFF or press the
CAL switch.
key for five seconds until
CC 0 .
Register the value and return to Step 2.
Cancel the settings without changing
the parameter and return to Step 2.
1CC
appears
57
CAL switch
CALスイッチ
Release the keys
Change the calibration
weight value if necessary
Weighing unit
Weighing unit
To output the results