Parameter
Item
Example
Example: Storing the unit mass of the counting mode
The unit mass is stored using10 samples.
Sign "-" means that something is placed on the weighing pan..
Example: Function settings
Select the item using the MODE key and enter it using the
ENTER key.
Input a parameter using the numerical keys.
Store the parameter using the ENTER key.
Example: While setting preset tare
PT
Input a tare value using the numerical keys.
Store the value using the ENTER key.
Example: Hold display
The hold display is set using Hold of the function table.
HOLD
When the weighing value is near zero (within the zero band) or
changes more than 25% +30 digits, the hold is canceled.
This error occurs when the mass sensor receives a strong force in
the upward direction.
Please check if the weighing pan is caught in something or
foreign matter is not entering the base part. There is a possibility
that the mass sensor or internal circuit has failed.
Overload display
Remove anything that is on the weighing pan.
Calibration Error
Please check if the weighing pan is properly placed or the weight
is too light and try again.
Calibration Error
Please check if the weighing pan is properly placed or the weight
is too heavy and try again.
This is an error where the zero value cannot be displayed when
the display is turned on. Please remove the items on the weighing
pan. If calibration is necessary, perform "Calibration of the Zero
Point" in "12.2. Calibration Using the Weight".
Or, this is an error where the weighing value is not stabilized
when the display is turned on. Avoid a breeze or vibration and
check whether the weighing pan is touching something.