Span Calibration
Step 5 The value of the calibration weight to be placed on the
weighing pan is displayed (The initial value depends on
the scale model).
The value for the calibration weight can be entered using
the numerical keypad.
To obtain proper weighing performance from the scale,
we recommend using a weight more than two-thirds the
weighing capacity.
Step 6
Place a weight of the value displayed on the pan.
Wait for the stability indicator to be displayed.
Press the ENTER key.
Step 7
Calibration using the weight is finished.
Cal appears after displaying
Remove the weight from the pan.
Step 8
Press the CAL switch to turn the display off.
Calibration using the weight is finished.
Step 9
Press the ON/OFF key to turn the display on.
Confirm that weighing can be performed properly.