AND HV-CWP Series Instruction Manual page 61

Waterproof digital platform scale
Hide thumbs Also See for HV-CWP Series:
Table of Contents

Advertisement

Output of "Title Block" and "End Block"
When weighing values are recorded as the GMP, GLP report, "Title
Block" and "End Block" are added at the beginning and at the end
of a group of weight values.
Title Block
Step 1
In the weighing mode, press and hold the PRINT key
5tart appears, and then release the key. The
until
scale outputs the "Title Block." The scale automatically
returns to the weighing mode.
Step 2
Press the PRINT key or select the auto-print mode to
output the weight values.
End Block
Step 3
Press and hold the PRINT key until
appears, and then release the key.
The scale outputs the "End Block."
Step 4
The scale automatically returns to the weighing mode.
recend1
59
ZERO
Press
and hold
Press
and hold
End block
To output the "
"

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents