Non-Zeeman Electrode Replacement - Agilent Technologies Graphite Tube Atomizer GTA 120 User Manual

Including gta viewing and fume extraction apparatus
Table of Contents

Advertisement

Maintenance

Non-Zeeman Electrode Replacement

Make sure you remove the electrode on the right before removing the left
NOTE
electrode.
The same procedure is relevant for the removal of both the left and right
electrodes, however the orientation of the two are reversed.
Powder free nitrile (or fabric) gloves should be used.
To replace the electrodes:
1
2
3
4
Ensure the electrode is straight, so that it goes into the housing aligned.
NOTE
Figure 24. Left electrode replacement
58
Open the furnace.
Put the adapter on the bolt, so the bolt head fits into the slot in
the adapter. This adapter will protect the electrode during
installation.
Being careful not to damage the electrode, place the electrode
over the bolt thread, thinner end first, until it sits in the adapter.
While holding the bolt and adapter onto the electrode, position
the electrode against the inner side of the electrode housing.
Agilent Graphite Tube Atomizer GTA 120 User's Guide