Zeeman Electrode Removal - Agilent Technologies Graphite Tube Atomizer GTA 120 User Manual

Including gta viewing and fume extraction apparatus
Table of Contents

Advertisement

5
6
To avoid cracking the electrode, do not over-tighten the nut. As soon as you feel
NOTE
resistance, stop tightening the nut.
7
8
9
10 After checking that it is clean, replace the center cooling block

Zeeman Electrode Removal

You will need:
Powder free nitrile (or fabric) gloves should be used.
NOTE
Agilent Graphite Tube Atomizer GTA 120 User's Guide
Put the pressure plate and PTFE washer on the bolt, and thread
the knurled nut onto the bolt. Using the wrench provided, screw
the nut onto the bolt until the assembly is held in position. The
shouldered face of the pressure plate fits into the window side of
the electrode housing.
Using the tool provided screw the nut until the electrode is
pulled squarely into the housing.
Unscrew the nut from the bolt and carefully remove the tooling.
Replace the window in the electrode housing.
Repeat Steps 1 to 8 for the right-hand electrode.
onto the left-hand electrode housing, being careful not to kink or
twist the tubes. Make sure neither tube is wedged between the
plates below the cooling block.
Zeeman electrode removal kit
Gloves
Set of new electrodes
Maintenance
59