Zeeman Gta - Agilent Technologies Graphite Tube Atomizer GTA 120 User Manual

Including gta viewing and fume extraction apparatus
Table of Contents

Advertisement

Using Your GTA
The umbilical should be recessed behind the front panel of the instrument. (The
NOTE
front panel has a cutout in it to accommodate this.)
7
8
For instruction about how to install and adjust the Programmable Sample
NOTE
Dispenser (PSD) refer to the PSD User's Guide.

Zeeman GTA

The furnace will have been installed by an Agilent field service
engineer during installation of the instrument. However, if you have
removed the furnace to replace the electrodes or clean the sample
compartment, reinstall the furnace workhead as follows:
1
2
Figure 4. Removing the left panel on the instrument
3
32
Align the furnace. Refer to the SpectrAA Help for details on how
to do this.
Replace the instrument left panel.
Ensure that both the spectrometer and GTA control unit are
switched off.
Remove the left panel on the instrument. To do this, lift the
bottom of the panel upward and then pull it forward (see
Figure 4).
Lift the furnace towards the sample compartment, ensuring the
umbilical cord is not twisted as you do so.
Agilent Graphite Tube Atomizer GTA 120 User's Guide