To remove the electrodes:
1
2
3
4
5
6
7
Insertion
1
Removal
6
Figure 27. Electrode replacement (right hand electrode shown) Zeeman
instrument
1. Draw bolt
2. Adaptor
3. Spacer
4. Stainless steel collar
8
Ensure that the spacer sits inside the rubber grommet and doesn't pinch the
NOTE
rubber when clamped.
Agilent Graphite Tube Atomizer GTA 120 User's Guide
Remove both window assemblies, by twisting and pulling.
Flip up the latch (Figure 26, item 1) on top of the furnace.
Open the furnace using the lever (Figure 26, item 5) and remove
the rubber skirt from the right side.
Loosen the magnet screw (Figure 26, item 2) a few turns.
Carefully pull the center block (Figure 26, item 3) of the furnace
upwards and clear of the workhead. Leave the tubing in place.
Lay out the components of the electrode replacement tool.
Pass the drawbolt through electrode on the right (from the
outside of the furnace). Refer to Figure 27.
5
7
2
7
3
4
5
Place the spacer over the electrode on the inside of the furnace.
6
1
2
5. PTFE washer
6. Draw nut assembly
7. Electrode pressure plate
Maintenance