Using Your GTA
Setting the Gas Supply
Turn on the gas supplies at the cylinders and set the regulators as
detailed in the 'Specifications' section on Page 18.
The use of pressures above the recommended maximum may burst internal gas
CAUTION
hoses, requiring service replacement.
Turning on the Water Supply
If you are using a recirculating system, you should refer to its
operating instructions.
Refer to the required water pressures listed in the 'Water Supply' section on
NOTE
Page 20.
If you are using a mains water supply, turn it on. The water pressure
regulator was set at the factory as per the specifications included in
Section 2.
The use of higher pressures may burst water hoses within the furnace assembly,
CAUTION
requiring service replacement.
After turning on your supply, you should check all hoses and
connections for leaks. If you find any you should contact your Agilent
field service engineer.
34
Agilent Graphite Tube Atomizer GTA 120 User's Guide