2.5 Ml/Min Flow Capillary Kit - Agilent Technologies 1100 Series System Manual

Capillary lc system
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

NOTE
.
capillary G1375-68703 between the column and the detector (item 8). Otherwise
change it as recommended with the capillary G1375-87308.
NOTE
.
Chapter

0.1 - 2.5 ml/min Flow Capillary Kit (5065-4495)

It is possible to use the capillary pump with a flow rate higher than 100 µl/min. For
this the pump must be used in the normal mode and some hardware modifications
are necessary.
From 100 to 200 µl/min, bypassing the electronic flow control is needed. No other
hardware modifications are necessary.
From 200 to 2500 µl/min bypassing the electronic flow sensor, installing the manual
purge valve (supplied with the accessory kit), changing the UV detector cell and
changing the capillaries in the flow path is needed.
The 0.1 to 2.5 ml/min flow capillary kit (5065-4495) include all the capillaries needed
to work with a flow from 200 to 2500 µl/min.
Table 81 0.1 - 2.5 ml/min flow capillary kit 5065-4495 content
Part number
G1375-87400
G1375-87318
G1375-87312
G1329-87302
G1375-87312
G1375-87306
G1316-87300
* SST: stainless steel
1100 Series Capillary LC System Manual
If the pressure drop in the system is not to high you can leave the
To successfully install the capillaries and avoid the risk of a leak go to
3, "Capillaries and Fittings.
Diameter
Pressure
(µm)
drop (Bar)
170
2
125
15
100
13
250
3
100
13
100
37
170
<1
133
Length
Material
(mm)
280
SST *
550
PFS **
200
PFS
1800
SST
200
PFS
550
PFS
70
SST
Options 6
Volume
Fitting
(µl)
type
6.4
A/A
6.8
B/C
1.6
B/C
88
B/B
1.6
B/C
4.4
C/B
1.6
A/A