Agilent Technologies 1100 Series System Manual page 56

Capillary lc system
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

3
Capillaries and Fittings
Item
Fitting
*
descriptions see
type
list below)
see description
A/A
1 below
see description
B/C
2
13
B/C
14 (micro ALS)
B/B
14 (micro WPS)
B/B
see description
B/C
3
see description
C/B
4
see description
A/A
5
11
A/A
* see
Table 14
on page 52
SST: Stainless steel
* * PFS: Peek coated fused silica
Descriptions
1 The capillary G1375-87400 connects the mixer and the manual purge valve.
2 The capillary G1375-87318 connects the manual purge valve and the injection valve (port1).
to
Table 10
3 The capillary G1375-87312 connects the injection valve (port 6) and the heat exchanger
4 The capillary G1375-87306 connects between the injection valve (port 6) and the heat
5 The capillary G1316-87300 connects between the heat exchanger (OUT) and the column.
The pressure drops in
NOTE
of
1000 µl/min.
Table 11 Specific capillaries for use with a micro CSV and a 20 µl flow sensor
46
1100 Series Capillary LC System Manual
Diameter
Length
(µm)
(mm)
170
280
125
550
100
200
250
1800
250
1800
100
200
100
550
170
70
170
380
(IN).
exchanger ( IN) if the thermostat G1330A/B is installed.
Table 10
Material
Volume
(µl)
SST
6.4
PFS**
6.8
PFS
1.6
SST
88
SST
88
PFS
1.6
PFS
4.4
SST
1.6
SST
8.6
are calculated for water (viscosity 1) and for a flow rate
Pressure
Part number
drop (Bar)
2
G1375-87400
15
G1375-87318
13
G1375-87312
3
G1329-87302
3
G1377-87300
13
G1375-87312
37
G1375-87306
<1
G1316-87300
3
G1315-87311