Agilent Technologies 1100 Series System Manual page 51

Capillary lc system
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

NOTE
The high flow rate used for Fast System Flush is not user-defined. For the Fast System Flush, the
NOTE:
pump automatically sets a predetermined maximum pressure limit. This pressure limit
is determined by the hardware configuration of the pump.
The flow rate used for Fast System Flush is the highest flow which can be delivered
without exceeding the pressure limit.
2 When the Fast System Flush time has elapsed, the micro well-plate sampler needle is
If multiple injections are in progress, the next injection will begin when Fast
Composition Change/Reconditioning is completed.
1100 Series Capillary LC System Manual
returned to the needle-seat. The pump returns to the normal operating mode,
reconditioning the column at the flow and initial composition defined in the current
method. The column is reconditioned for the Column Reconditioning time defined in
the user interface.
41
Optimizing Performance 2