Never use cables other than the ones supplied by Agilent Technologies to ensure proper
functionality and compliance with safety or EMC regulations.
Bench Space
The modules dimensions and weight (see
Capillary LC system on almost any laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in
the rear for the circulation of air and electric connections.
If a Thermostatted Well Plate Sampler is installed, an additional 25 cm (10 inches)
of space on either side for the circulation of air, and approximately 8 cm (3.1
WARNING
inches) at the rear for electrical connections is required.
If a complete Agilent Capillary LC system is to be installed on the bench, make sure that
the bench is designed to carry the weight of all the modules. For a complete system
including the Thermostatted Well Plate Sampler it is recommended to position the
modules in two stacks. see
Environment
Your modules will work within specifications at ambient temperatures and relative
humidity as described in
ASTM drift tests require a temperature change below 2°C/hour (3.6 °F/hour) over one
hour period. Our published drift specification (refer also to
Agilent 1100 Series Capillary Pump"
ambient temperature changes will result in larger drift.
Better drift performance depends on better control of the temperature fluctuations. To
realize the highest performance, minimize the frequency and the amplitude of the
temperature changes to below 1 °C/hour (1.8 °F/hour). Turbulences around one minute
or less can be ignored.
Do not store, ship or use your modules under conditions where temperature
CAUTION
fluctuations could cause condensation within the modules. Condensation will damage
the system electronics. If your modules were shipped in cold weather, leave them in their boxes and
allow them to warm slowly to room temperature to avoid condensation.

Physical Specifications

1100 Series Capillary LC System Manual
Table 2
"System Installation Process"
Table 1
on page 4.
on page 166) is based on these conditions. Larger
3
Installing your Capillary LC System 1
on page 5) allow to place the
on page 6.