Checking for Leaks

Leaks in Capillary Flow Fittings

WA R N I N G
104
For capillary flow fittings, a leak usually indicates that the
fitting has been overtightened. Unless the fitting is obviously
loose, do not tighten it further. Instead, remove the connection,
trim the column end, and install it again. (See
Capillary Column Using SilTite Metal Fittings"
Also inspect the plate and connection for a broken column tip.
Be careful! The LTM column, inlet, guard column heated
enclosure, GC/MSD transfer line may be hot enough to cause
burns. If hot, wear heat-resistant gloves to protect your hands.
"To Attach a
on page 114)