Agilent Technologies 1100 Series System Manual page 175

Capillary lc system
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Detection type
Light source
Wavelength range
Short term noise
(ASTM)
Single and
Multi-Wavelength
Drift
Linear absorbance
range
Wavelength accuracy ± 1 nm
Wavelength
bunching
Slit width
Diode width
Flow cell
Table 92 Performance Specifications Agilent 1100 Series Diode Array Detector
Type
Specification
Maximum pressure
50 bar
Control and data
Agilent ChemStation for LC
evaluation
Analog outputs
Recorder/integrator: 100 mV or 1 V, output
range 0.001 – 2 AU, two outputs
1100 Series Capillary LC System Manual
1024- element photodiode array
Deuterium and tungsten lamps
190 – 950 nm
Typically ± 3 × 10
rates <100 µl/min
*
-3
2 × 10
AU/hr at 254 nm
> 2 AU (upper limit)
1 – 400 nm
1, 2, 4 , 8, 16 nm
< 1 nm
500 nanoliter: 0.5 µl volume, 10 mm cell
path length and 50 bar (725 psi) pressure
maximum
165
Performance Specifications
-5
AU at 254 nm at flow
Comments
For the 500 nl flow cell
the noise is 2-3 times
higher
than
with
standard flow cell
Self-calibration with
deuterium lines,
verification with
holmium oxide filter
Programmable in steps
of
1 nm
Programmable slit
7