9.0 Specifications

Voltage range
Current input
Power cable
Power consumption
Operating modes
Stimulation time
Timer indication
Air temperature
Lowest temperature
Temperature indication
Temperature deviation
Air flow
Connections
Operating time
Fuses
Protective earth conductor resistance
Earth leakage current
Enclosure leakage current
Patient leakage current
Ambient conditions
Transport/storage
Operation
Dimensions HxWxD
Weight
Protection class (EN 60601-1)
Degree of protection
Protection category
Classification acc. to Annex IX
EEC directions 93/42/EEC
CE marking
Rules applied
UMDNS-Code
REF
Canadian Classification
Device group
PNC
Risk Class
Description
42
UK
100 - 240 V~ ± 5 %; 50/60 Hz
0.75 A max.
3 m
85 W max.
Preparation mode; Stimulation mode (at temperature preselected for the cold
stimulation level resp. warm stimulation level); Heating off and no air flow
(economy mode, standby mode)
can be preselected by timer from 1 up to 99 sec.
Accuracy ± 0.5 s ± ½ digit
20°C - 47°C
approx. 2°C above room temperature
Accuracy ± 0.5 °C ± ½ digit
< ± 1°C
5.0 l/min ± 10 %
Power connection; Control output for a nystagmograph; Equipotential
equalization; Connection for the handle; Air inlet
Short-time operation:
1. Automatic shut-off after completion of
stimulation.
2. Automatic shut-off after 3 minutes.
T 0,8 A
0.1 Ω max.
0.5 mA max.
0.1 mA max.
0.1 mA max.
-20...+50°C; 5...90 % humidity without condensing,
air pressure 700...1060 hPa
+10...+35°C; 20...80 % humidity without condensing,
air pressure 700...1060 hPa
14.5 x 37 x 32 cm
approx. 3.7 kg
I
Type B
IPX0
IIa
CE 0124
EN 60601-1: 1990 + A1:1993 + A2:1995
EN 60601-1-2: 1993 (EMV / EMC)
10-548
502.1100.0
Ear, nose, throat
77ETP
2
STIMULATOR, CALORIC (WATER)
F
Issue of the Technical Specifications: 01.04.2005
Page 43