Visual Inspection; Electrical Test; Replacing Of Filter - Atmos Varioair 3 Operating Instructions Manual

Hide thumbs Also See for Varioair 3:
Table of Contents

Advertisement

6.2

Visual inspection

Test
Check the unit for external damage (damage to cables, hoses, filter and handle).
Check filter for dirt or a defect.
Inspect unit, disconnected from the main power supply and with open housing, for internal damage, and leaks.
Close housing again and tighten screws. Re-connect the main power cable.
Check fuses for correct protection value and melting characteristics.
Check sign plates and inscriptions for completeness and readability.
6.3

Electrical test

Test
Measure protective conductor resistance according to VDE 0751 (max. 0.3 Ohm).
Measure and compare with "first value measured" of unit leakage current according to VDE 0751
(measured during operation).
The "value measured first" may not be exceeded by more than 50%.
Document and record the evaluation and tests according to VDE 0751.
6.4
Replacing the filter
The air filter (
, fig. 2) must be replaced once per year as follows:
• Remove the filter by unscrewing it from the holder (turn the filter to the left).
• Mount the new filter by screwing it into the holder (turn the filter to the right).