11.0 Notes on EMC

Immunity Test
Voltage Dips / Dropout
IEC 61000-4-11
NOTE
U
is the mains alternating current prior to application of the test levels.
T
11.3 Guidelines and Manufacturer´s Declaration - Immunity
The ATMOS Varioair 3 is intended for use in the electromagnetic environment specified below. The customer or user
of the ATMOS Varioair 3 should ensure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601-
Test Level
IEC 60601-
Compliance Level
Test Level
3 V
[3] V
eff
150 kHz to 80 MHz
3 V/m
[3] V/m
80 MHz to 2.5 GHz
Compliance Level
< 5 % U
T
(> 95 % Dip of the U
)
T
for 0.5 Cycle
40 % U
T
(60% Dip of the U
)
T
for 5 Cycles
70% U
T
(30 % Dip of the U
)
T
for 25 Cycles
< 5 % U
T
(>95 % Dip of the U
)
T
for 5 s
Electromagnetic Environment
- Guidance
Portable and mobile communications
equipment should be separated from the
ATMOS Varioair 3 incl. the cables by no
less than the distances calculated/listed
below.
Recommended distances:
d = [ 1,2] √P
d = [ 1,2] √P
d = [ 2,3] √P
where „P" is the max. power in watts (W)
and D is the recommended separation
distance in meters (m).
Field strengths from fixed transmitters, as
determined by an electromagnetic site (a)
survey, should be less than the compli-
ance level (b).
Interference may occur in the vicinity of
equipment containing following symbol.
Electromagnetic Environ-
ment - Guidance
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the Vari-
oair 3 demands continued function
even in case of interruptions of the
energy supply, it is recommended
to supply the
ATMOS Varioair 3 from an
uninterruptible current supply or a
battery.
UK