10.0 Technical specifications
Air flow rate of pump
Max. vacuum
Vacuum readout
Additional air regulation
Collection jar
Suction hose
Voltage
Current input (max.)
Operating time
Fuse
Heat emission
Noise level
Ambient conditions
Transport/storage
Operation
Dimensions HxWxD
Weight
Regular safety relevant
inspections
Protection class (EN 60601-1)
Degree of protection
Protection category
Classification acc. to Annex IXEEC
directions 93/42/EEC
CE marking
Rules applied
UMDNS-Code
GMDN-Code
REF
* 1 bar = 750,06 mm Hg =1000 hPa / depending on ambient air pressure
45 l/min +3/-5 l/min
-91kPa ( -910 mbar oder 682,5 mmHg)*@ NN
-1...0 bar ± 16 mbar (Klasse 1,6) ø 63 mm
mechanical regulating valve
1.5 l, 3 l or 5 l jars DDS or Receptal
3 l, with use of the trolley
ø 6 mm or ø 10 mm
230 V~ (+/-10%) 50/60 Hz
Special voltage:
115 V ~ (+/-10%) 60 Hz
max. 0,75 A bei 230 V~
max. 1,3 A bei 115 V~
> 12 h continuous operation without interruption within 24 h
T 800 mA/H for 230 V~
T 1,6 A/H for 100 - 127 V~
max. 173 J/s
≤ 48 dB (A) @ 1 m (ISO 7779) bei max. Vakuum
-30...+50°C
5...90 % humidity,
non-condensing air pressure 700...1060 hPa
+5...+35°C
20...80 % humidity,
non-condensing air pressure 700...1060 hPa
H 330 x B 240 x T 360 mm (with secretion container)
H 1010 x B 360 x T 440 mm (with system trolley)
(with collection jar)
Ca. 6,7 kg
No
II
Typ BF
IPX 1
IIa (acc. to EC directive 93/42 EEC)
CE 0124
EN 60601-1: 2007
EN ISO 10079-1: 2000
10 - 217
36777
340.0300.0
230 V
340.0300.2
115 V
®
- system 1.5 l, 2 l,
Technical data issued on: 04.07.2011