Stator Status, Led B (Lower Led) - HBM T40FM Mounting Instructions

Torque flange
Hide thumbs Also See for T40FM:
Table of Contents

Advertisement

Available languages

Available languages

T40FM

7.2 Stator status, LED B (lower LED)

Color
Green
(permanently lit)
Green, intermittently
orange.
Numerous
synchronization
defects:
permanently orange
Orange
(permanently lit)
Red
(permanently lit)
A3276-3.0 en/de
Meaning
Measurement signal transmission and internal stator voltages ok
Orange until end of defective transmission if y5 incorrect
measured values in sequence are transmitted. The
measurement signals reflect the level of the defect status for
the duration of the transmission defect + for approx. another
3.3 ms.
Permanently disrupted transmission, the measurement
signals reflect the level of the defect status. (f
= defect level).
=> Correct the rotor/stator alignment.
Internal stator defect, the measurement signals reflect the
level of the defect status (f
out
= 0 Hz, U
= defect level).
out
out
33
= 0 Hz, U
out
HBM