Abbott ARCHITECT RHS Training Manual page 95

Retest sample handler
Table of Contents

Advertisement

Procedure
5501 RSH Motor Test
5506 RSH Sensor Test
Sample Probe Calibration
1120 Sample Pipettor
®
Calibration (c8000
System)
1111 Sample Pipettor
®
Calibration (i2000
System)
SR
1117 STAT Pipettor Calibration
®
(i2000
System)
SR
Sample Pipettor Check
2043 Sample Pipettor Check
®
(c8000
System)
2001 Sample Pipettor Check
®
(i2000
System)
SR
®
ARCHITECT
RSH System Service Training Guide
201532-102
FUNCTION and USE
M or D
D
• Used to home the following RSH motors:
carrier transport X, carrier transport theta,
carrier transport Z, and carrier positioner.
D
• Tests the carrier transport and carrier
positioner sensors.
M
• Automatically determines the position of
calibration targets in relation to the home
position of the pipettor.
• Probe Straightness is determined.
• Positions are set for Sample Carriers, or
LAS Sample Carousel Wash station, RV1
and RV24.
• STAT positions are set for Sample
Carriers, Wash Station, and RV 47
M
• Precision check which compares the
amount of saline dispensed by the Sample
Pipettor to a reference
• Precision check which compares the
amount of saline dispensed by the Sample
Pipettor to a reference
MODULE 3: ROBOTICS
• Troubleshooting
• Troubleshooting
• Install
• Install
• Troubleshooting
• Component Replacement
• Troubleshooting
.
.
FOR INTERNAL USE ONLY
When to Perform
3-21