Abbott ARCHITECT RHS Training Manual page 124

Retest sample handler
Table of Contents

Advertisement

MODULE 4: POWER AND CIRCUIT BOARDS
TROUBLESHOOTING
The following tips are for troubleshooting Power System failures:
NOTE: When troubleshooting RSH System errors refer to your ARCHITECT RSH
Symptom
RSH OFFLINE or will not boot-up
Motor moves freely with power ON
Motor Encoder Errors
5417 Section Align Sensor Failure
Section of indicator lights on after
installation at Ready Status
4-22
FOR INTERNAL USE ONLY
System Service and Support Manual, On-line Troubleshooting information
(eSolutions), Technical Service Bulletins, Instrument Service Advisories, and/
or the ARCHITECT System Operations Manual.
Possible cause of failure
• Power is OFF
• SMD/SH PS5 or PS6 failure
• Cables Loose on RSH
• Ethernet cable not connected at PM or HUB
• Loss of 36VDC
• Loose cable on Card Cage or Assembly
• Motor Driver failure
• Indexer Board failure
• Damaged or loose connection at assembly
• Failed Positioner Sensor
• Damaged Carrier Transport cable sensor or Flex Board
• Dirty Carrier Transport Section Sensor
• Loose cable on Carrier Transport, SIB or Card Cage
• Fuse F2 on SIB blown
• RSH Controller Board failure
• UDB switch incorrectly set
• SIB, UDB, PLSB failure
• UDB or PLSB not connected correctly to next board
®