REVIEW QUESTIONS AND CASE STUDIES
SCENARIO 1 There have been several occurrences of error 5417 Section Align Sensor Failure error. The
operator noted the error started occurring during RUNNING status. Now the RSH is down and the
customer cannot use it without this error occurring.
The following troubleshooting STEPs were performed, but the error remains unresolved:
• 1119 Transport Calibration
• Cleaned and checked transport and sensors
• Reseated transport cables
What checks, measurements, procedures, etc. would you use to help isolate the root cause of the
failure? What would you do to try and resolve this error? What do you think may be causing the
failure?
SCENARIO 2 Error 5417 (Carrier Positioner 2 encoder) sensor failure on a
were detected.
The following troubleshooting STEPs were performed, but the error remains unresolved:
• 1119 Transport Calibration
• Sensor Test
• Reseated SIB cables
What checks, measurements, procedures, etc. would you use to help isolate the root cause of the
failure? What would you do to try and resolve this error? What do you think may be causing the
failure?
®
ARCHITECT
RSH Service Training Guide
201532-102
MODULE 4: POWER AND CIRCUIT BOARDS
®
ci8200
System. No obstructions
FOR INTERNAL USE ONLY
4-29