REVIEW QUESTIONS AND CASE STUDIES
SCENARIO 1 5417 (Section Align) sensor failure errors are occurring on a ci8200 System. This has been a
reoccurring problem. This is the 5th time this has been reported within the last 6 weeks.
On a prior visit to repair this error the Carrier Transport Assembly was replaced for the second
time.
Currently, the following troubleshooting STEPs have been performed and the issue remains
unresolved:
• Cables to Carrier Transport have been reseated
• SIB voltages are OK
• Card Cage voltages are OK
What checks, measurements, procedures, etc. would you use to help isolate the root cause of the
failure? What would you do to try and resolve this error? What do you think may be causing the
failure?
1.
In a multi-module system, which Processing Module controls the Sample Handler (RSH) power?
2. What type of errors could occur if the RSH and Process Modules are not properly leveled?
3. The Priority Bay is placed on which end of an ARCHITECT ci8200 RSH?
®
ARCHITECT
RSH System Service Training Guide
201532-102
MODULE 5: MISCELLANEOUS
FOR INTERNAL USE ONLY
5-21