Abbott ARCHITECT RHS Training Manual page 92

Retest sample handler
Table of Contents

Advertisement

MODULE 3: ROBOTICS
Diagnostic Information:
Carrier Positioner Board receives +5VDC from SIB
Carrier Positioner Board provides +5VDC to the Carrier Positioner Encoder Sensor
Encoder Sensor is replaceable
Carrier Positioner encoder may be mounted at an angle causing 5417 encoder sensor failures
to occur.
5409 Step Loss errors can be caused by:
-
Loose Lead Screw nut on Carrier Positioner Drive Assembly
-
Loose 1.5mm drive set screws on the Carrier Positioner Drive Assembly motor coupler
-
Loose Lead Screw on Carriage Block
5409 homing failures can be caused by:
-
Loose Lead Screw on Carriage Block
-
Dirt, debris, obstruction
3-18
FOR INTERNAL USE ONLY
GOOD
BAD
Coupler
Set Screws
Carriage Block with
set screws
®
ARCHITECT
RSH Service Training Guide
201532-102

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents