4.
Prepare a waste basket.
Figure 19 Waste basket
5.
Turn on the AC power switch of the Peltier Temperature Controller 201.
Figure 20 Turn on the AC power switch
6.
Connect the waste hose to the Coolant Drain port on the rear panel.
Figure 21 Connect the waste hose
The Coolant flows out as soon as the waste hose is connected, so the other side of the waste hose should
NOTE:
be placed into the waste basket before connected.
7.
The coolant will be drained automatically.