a. Y Unit: Select one of the display unit : Absorbance, Transmittance, Reflectance or Energy.
b. SBW (nm): Select bandwidth. There are five bandwidths selectable: 0.5, 1, 2, 5 or 20 nm.
c. Tm Method: Select a method for determining melting temperature. Options include: 1st
derivative and Average.
d. Equation Name: Enter the name of the formula. %G-C is entered as default, of which
equation is formulated in Equation Expression.
e. Equation Expression: The default equation for the calculation of %G-C base pairs is:
%G-C=2.44 * (Tm - 81.5 - 16.66 * log(M) + 500/K)
Where M is the molarities of salt in mol/L, K is the DNA base pair length. %G-C will be
calculated with estimated T
f. DNA Pair Length(K)
Enter the DNA base pair length. This value will be used calculating %G-C.
g. Calculation Start Temperature (°C): Enter the start temperature for calculating the T
value.
h. Calculation End Temperature (°C): Enter the end temperature for calculating the T
i.
Molarity (mol/L): Enter the salt molarities of the solution [mol/L]. This value will be used in
calculating %G-C.
7.
After setting parameters for Experiment and Lamp, click the Accessory tab. And select Multi-
Cell Peltier.
and input values, Molarities (M) and length of DNA (K).