Installing/removing modules in the rack-mounted enclosure

5.4 Installing/removing analyzer modules

5.4
Installing/removing analyzer modules
5.4.1

Note on installation / connecting up

CAN bus cable
NOTICE
Confusing the connectors when connecting the modules
When connecting the modules to the CAN bus, confusing the connectors can lead to damage
to the device.
● For this reason, note the assignment shown in the following figure.
● Orient yourself in particular on the alignment of the CAN bus coding mark as of which the
OM2.2
PU
AM1/
AM2
nc
Figure 5-2
70
connectors are coded differently.
CAN bus connectors
CAN bus cable
Coding mark
Connectors of option module 2.2
Replacing the processing module
Connectors for the analyzer modules 1 and 2
Free no function
Schematic representation of the CAN bus cable
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03