5.4.2
Installation of the analyzer module
Safety instruction(s)
NOTICE
Installation of ULTRAMAT 7 analyzer modules
ULTRAMAT 7 analyzer modules-that are installed contrary to their intended purpose in
another device variant can cause damage. Therefore:
● Prior to installation, check the device variant for which the analyzer module is intended.
● Install ULTRAMAT 7 analyzer modules manufactured for operation in a slide-in device
● Install ULTRAMAT 7 analyzer modules manufactured for operation in a wall device only
You can obtain information about this from your Siemens partner→ Technical support
(Page 144)..
Procedure
①
Figure 5-3
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03
You will find the relevant information in the MLFB number on the module plate:
– 7MB3010-xxxxx-xxAx-Z → "A": Analyzer module for slide-in devices
– 7MB3010-xxxxx-xxBx-Z → "B": Analyzer module for wall devices
only in slide-in devices.
in wall devices.
Rack-mounted enclosure
Installing an analyzer module based on the example of an OXYMAT 7
Installing/removing modules in the rack-mounted enclosure
5.4 Installing/removing analyzer modules
②
OXYMAT 7 analyzer module
71