Technical Data Ultramat 7 - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis, rack-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Table 11-14 OXYMAT 7: Gas connections
Gas connections
With hoses
With pipes
Table 11-15 OXYMAT 7: Parts in contact with the sample gas
Materials in contact with the sample gas
Sample chamber
Gas path
Sealing material
Special applications
11.3.2

Technical data ULTRAMAT 7

Table 11-16 ULTRAMAT 7: General technical specifications
General
Information on the smallest span
Power consumption
Weight
Electrical safety
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03
11.3 Technical specifications of analyzer modules
Plastic screw connection for plastic pipe or hose 4 mm / 6 mm
Connection fittings for 6 mm pipe
Stainless steel:
● Plates: Mat. no. 1.4571 (X6CrNiMoTi17-12-2)
● Screw-in glands: Mat. no. 1.4404 (X2CrNiMo17-12-2)
Hastelloy:
● Plates: Mat. no. 2.4602 (NiCr21Mo14W)
● Screw-in glands: Mat. no. 2.4819 (NiMo16Cr15W)
With hoses
FPM (e.g. Viton), connections PVDF
With pipes
Stainless steel:
● Pipes: Mat. no. 1.4571 (X6CrNiMoTi17-12-2)
● Gas connections: Mat. no. 1.4571 (X6CrNiMoTi17-12-2)
Hastelloy:
● Pipes: Mat. no. 2.4602 (NiCr21Mo14W)
● Gas connections: Mat. no. 2.4819 (NiMo16Cr15W)
FPM (e.g. Viton) or FFKM (Kalrez 6375)
Pipe gas path
Materials adapted to the application
See module nameplate
≤ 30 W
Max. 5.2 kg (standard version)
In accordance with DIN EN 61010-1
Technical specifications