Advertisement

Installing/removing modules in the rack-mounted enclosure
5.1 Safety instructions
Signal frequencies
Note
Signal frequencies for operation with two OXYMAT 7 analyzer modules
If you install a second OXYMAT 7 analyzer module, the operating and signal frequencies of
the two OXYMAT 7 analyzer modules have to differ from each other.
● Check the set frequencies of both OXYMAT 7 analyzer modules.
● If necessary, adjust the frequencies of the modules. We recommend a frequency of 8.33
Hz for Module 1 and a frequency of 10.00 Hz for Module 2.
You can find additional information in the bibliography in:
→ Table A-3 References 1 - Operating Manuals LUI (Page 141)
→ Table A-4 References 2 - Operating Manuals PDM (Page 141)
Note
Combined operation with ULTRAMAT 7: Signal frequency of the OXYMAT 7
If you use an OXYMAT 7 and an ULTRAMAT 7 analyzer module in combined operation, the
default OXYMAT 7 signal frequency can have a negative influence on the ULTRAMAT 7
analyzer module.
● Check the set frequency of the OXYMAT 7 analyzer module.
● If the nominal frequency 8.33 Hz is still set, change the signal frequency to 10 Hz.
You will find additional information in the list of references in:
→ Table A-3 References 1 - Operating Manuals LUI (Page 141)
→ Table A-4 References 2 - Operating Manuals PDM (Page 141)
See also
References (Page 141)
66
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03