Installing the analyzer modules and option module
5.1
Tools
Designation
Screwdriver
Hexagon socket wrench
Torque wrench 0.2 to 2.5 Nm
Open-ended wrench SW7
5.2
Safety notes for installation
Damage to membrane keyboard and display
Make sure that the membrane keyboard and display are not damaged during installation.
CAUTION
Improper installation or modification can damage the device and thus invalidate the
approval.
Hot surface
Following shutdown, the analyzer module is still very hot. Wait until the analyzer module
has cooled down before dismounting.
NOTICE
Opening the rack-mounted housing is only permitted for installation or maintenance.
NOTICE
If the device is partially or completely open, the degree of protection specified in the
technical data is no longer guaranteed.
Rack-mounted device
Operating Instructions, 01/2013, A5E31874006-01
Torx T20
Torx T10
SW7
SW5
CAUTION
CAUTION
Quantity
1
1
1
1
1
1
5
27