Calomat 7 Technical Specifications - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis, rack-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
11.3 Technical specifications of analyzer modules
11.3.3

CALOMAT 7 technical specifications

Table 11-26 CALOMAT 7: General technical specifications
General
Information on the smallest span
Power consumption
Weight
Electrical safety
Table 11-27 CALOMAT 7: Measuring ranges
Measuring ranges
Number of measuring ranges
Parameters can be assigned in the
measuring ranges
Table 11-28 CALOMAT 7: Gas inlet conditions
Gas inlet conditions
Sample gas pressure
Pressure loss between sample gas inlet
and sample gas outlet
Sample gas flow
Sample gas temperature and humidity
Table 11-29 CALOMAT 7: Temperature of sample chamber and gas path
Temperature of sample chamber and gas path
Gas path
Standard version
Table 11-30 CALOMAT 7: Time response
Time response
Warm-up period at room temperature
Response characteristics
132
Smallest possible span
Largest possible span
Smallest possible span with sup‐
pressed zero point
See module nameplate
< 20 W
Approx. 3 kg
In accordance with DIN EN 61010-1
Max. 4; parameters can be assigned freely
0.5% H
in N
2
2
100% H
in N
2
2
5% (e.g. 95% to 100%) H
700 to 1 200 hPa (absolute)
< 50 hPa at 1.5 l/min
30 to 90 l/h (0.5 to 1.5 l/min)
0 to 70 °C with relative humidity < 90 %, avoid condensation
in the gas path (moisture content results in measurement
errors)
5 K above ambient temperature
72 °C
< 30 min (max. accuracy after 2 h)
Operating Instructions, 05/2015, A5E31874006-03
in N
2
2
Rack-mounted device