Operation with at least one OXYMAT 7

Measures
The sample gas flow is created by overpressure of the process gas to be analyzed.
Pressure surges are created, which can lead to an unsteady measured value display:
● If the sample gas flow is subjected to rapid fluctuations in pressure
● When the sample gas is fed to the analyzer module by a pump.
The following measures protect the device from pressure surges:
● Installation of the supplied sample gas restrictor at the sample gas inlet of the OXYMAT 7
● Installation of a "pneumatic low pass" consisting of a restrictor and a damping vessel
Installation conditions for the sample gas restrictor
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03
Sample gas restrictor for OXYMAT 7, ver‐
sion with hoses
Rack-mounted device, rear view
4.2 Operation with at least one OXYMAT 7
Sample gas restrictor for OXYMAT 7, ver‐
sion with pipes
Application planning